4.2       Posology and method of administration
Prior to administration appropriate resuscitation equipment should be available.
DaTSCAN should only be used in adult patients referred by physicians experienced in the management of movement disorders and/or dementia. DaTSCAN should only be used by qualified personnel with the appropriate government authorisation for the use and manipulation of radionuclides within a designated clinical setting.
Posology
Clinical efficacy has been demonstrated across the range 111 to 185 MBq. Do not exceed 185 MBq and do not use when the activity is below 110 MBq.
Patients must undergo appropriate thyroid blocking treatment prior to injection to minimise thyroid uptake of radioactive iodine, for example by oral administration of approximately 120 mg potassium iodide 1 to 4 hours prior to injection of DaTSCAN.



1
Special populations
Renal and hepatic impairment
Formal studies have not been carried out in patients with significant renal or hepatic impairment. No data are available (see section 4.4).
Paediatric population
The safety and efficacy of DaTSCAN in children aged 0 to 18 years has not been established. No data are available.

Method of Administration
For intravenous use.
DaTSCAN should be used without dilution. To minimise the potential for pain at the injection site during administration, a slow intravenous injection (not less than 15 to 20 seconds) via an arm vein is recommended.
Image acquisition
SPECT imaging should take place between three and six hours post-injection. Images should be acquired using a gamma camera fitted with a high-resolution collimator and calibrated using the 159 keV photopeak and a ± 10% energy window. Angular sampling should preferably be not less than 120 views over 360 degrees. For high resolution collimators the radius of rotation should be consistent and set as small as possible (typically 11-15 cm). Experimental studies with a striatal phantom, suggest that optimal images are obtained with matrix size and zoom factors selected to give a pixel size of 3.5 – 4.5 mm for those systems currently in use. A minimum of 500k counts should be collected for optimal images.

שימוש לפי פנקס קופ''ח כללית 1994 לא צוין
תאריך הכללה מקורי בסל לא צוין
הגבלות לא צוין

רישום

131 95 30983 00

מחיר

0 ₪

מידע נוסף

עלון מידע לרופא

16.12.21 - עלון לרופא 21.07.24 - עלון לרופא

עלון מידע לצרכן

08.12.13 - עלון לצרכן 16.12.21 - החמרה לעלון 21.07.24 - החמרה לעלון

לתרופה במאגר משרד הבריאות

דטסקאן

קישורים נוספים

RxList WebMD Drugs.com